The EPOC and COAG Analyzers will need to have the Patient ID or Request ID MANUALLY entered into the analyzers as they currently will not pickup requests from the worklist, but will send the results back to Instinct.
Workflow
Complete the treatment in Instinct
Enter the Patient ID or Request ID into the analyzer
Run the sample
Results will be sent back to Instinct automatically
Best Practices for Manual ID Entry
The Request ID will display at the top of the Chart pane:
EPOC
The numeric portion of this number (after DR) can be entered in either the ‘Pat ID’ or ‘ID2’ field.
COAG
Enter the Requisition ID into the Patient ID or Sample ID field
Laboratory Treatments – Chart Screen
When a laboratory treatment is completed, a new Diagnostic Chart document is automatically created for the current visit.
The status of the Chart document will either say Pending if results are not yet completed, or Complete if results are returned (visible in real time).